Quantitative Imaging and Fluorescence Microscopy: Towards quantitative fluorescence microscopy: A new solution for standardization, monitoring, and quality management

June 1, 2015
Quantitative fluorescence microscopy is beginning to open new research opportunities in biosciences.  

ARNAUD ROYON

Fluorescence microscopes have become ubiquitous in life sciences laboratories, including those focused on pharmaceuticals, diagnosis, and forensics. For the past few years, the need for both performance guarantees and quantifiable results has driven development in this area.1 However, the lack of appropriate standards and reference materials makes it difficult or impossible to compare the results of two fluorescence microscopes, or to measure performance fluctuations of one microscope over time. Therefore, the operation of fluorescence microscopes is not monitored as often as their use warrants—an issue that is recognized by both systems manufacturers and national metrology institutes.2,3

This, of course, calls into question the reliability of collected data. Performance fluctuations of fluorescence imaging systems can impact several measurement types, including detection of investigated elements and measurement of their shape, dimensions, motion, and intensity.

According to the broad community of fluorescence microscopists, the ideal calibration tool for fluorescence microscopy would characterize (non-exhaustively) point-spread function; spatial resolution (x, y, and z); spectral irradiance reaching the sample; illumination homogeneity; field flatness; z-distance; spectral resolution; day-to-day and long-term instrument performance; and the detector's range of linearity, dynamic, and limit of detection.3 It should also not undergo photobleaching, and be reliable, durable, and easy to use and handle.

Sub-micrometer structured fluorescence

A process developed by Argolight, a start-up from the University of Bordeaux (France), was designed to address these needs. It marks and induces multi-scale (from sub-micron to centimeter), stable fluorescent objects in three dimensions (3D) inside transparent materials, ideally enabling the assessment of widefield, confocal, and multiphoton fluorescence microscopes. Figure 1a shows a typical multi-dimensional (space, intensity, spectrum, lifetime) calibration slide, consisting of a glass substrate on a stainless steel carrier. It features the same dimensions as a standard microscope slide. Various patterns are embedded in the glass, each one intended and designed to respond to a particular misalignment issue or failure of the system (see Fig. 1b).

These slides are manufactured from custom photosensitive zinc-phosphate glass. An ultrafast laser, emitting 200 fs pulses at 1030 nm with an average power of 5 W, locally modifies the chemistry of the glass, inducing such fluorescence patterns as 2D and 3D structures, straight and curved lines, dots, "stairs," and even double helixes (see Fig. 2). Fluorescence emitters embedded in the glass matrix are extremely stable, even in harsh environments.
The spectral properties of these structures include excitation ranging from 325 to 650 nm (see Fig. 3a), emission of a continuum up to 800 nm (see Fig. 3b), sub-nanosecond lifetime (see Fig. 3c), and absence of photobleaching. Adjustment of laser "dosage" enables control of emission intensity. These characteristics make the tool suitable for both spectral (see Fig. 3d) and intensity (see Fig. 3e) investigations of fluorescence-based systems.
Every such slide emits the same quantity of fluorescence for the same excitation conditions no matter when it is used, making it possible, for the first time, to quantify luminosity at pertinent spatial scales over extended periods (the slides are warranted for five years). This unique feature, which is specific to this technology, enables quick and complete characterization of several fluorescence microscopes, and monitoring of their performance over time—with reasonable monetary and time investments. Proprietary image analysis software further enables quality management (see Fig. 4). Because it is possible to compare results obtained by different researchers working in disparate locations, it also facilitates collaborative work.

The ability to obtain quantitative measurements from collected images enables dramatic improvements. This is true wherever the technology is used, but particularly in the pharmaceutical industry, where it enables, for example, following the effects of a drug over extended periods, and in medicine, where it improves the reliability of fluorescence-based diagnosis.

Measuring microscopes

Researchers at the Friedrich Miescher Institute for Biomedical Research (Basel, Switzerland) monitored the intensity response of a widefield fluorescence microscope for nearly five months using an Argolight calibration slide (see Fig. 5). The research team has measured fluctuations of ±15%, clearly illustrating changes in the intensity response of the device, from either its illumination components (light source, filter) or its detection assembly (camera, filter). For the cell or developmental biologist who wants to perform a study over time, if the intensity of the images is meant to be exploited, it is necessary to normalize them with respect to the intensity response of the system. This example illustrates measurement of just one aspect; additional parameters may also vary.

Intensity fluctuations over time in confocal microscopes can be much larger. In fact, the type of instrument is one factor in determining how often microscopes should be measured. Other variables that weigh into this decision include components of the microscope (mercury lamp vs. more stable LEDs in a widefield microscope), experience of the user (an experienced microscopist will notice changes more readily than a novice), whether the system is shared, the type of experiment for which it is used, and the level of quality needed. For example, confocal microscopes tend to fluctuate more than widefield microscopes simply because they have more components (such as piezo stages, galvo mirrors, lasers, and photomultiplier tubes) that are subject to change. And a researcher making a single measurement would not need to measure performance but simply know the system performs well, whereas one who needs to make multiple measurements at different times would have to know the fluctuations to compare measurements. Some pharmaceutical companies require users to calibrate their systems once a day.

With this technology, it takes about 15 minutes to monitor the performances of a microscope, compared with two hours using a combination of other technologies.

For the future

The standard Argo-M slide covers a vast range of microscopes, but does not fill the needs of all microscopists—so custom Argo-U slides can be tailored to meet virtually any requirements. And four new types of standard slides will become available in September 2015: the Argo-HM (High Magnification) for magnifications ≥40x, the Argo-LM (Low Magnification) for magnifications ≤20x, the Argo-SIM for structured illumination microscopes, and the Argo-HCS for high content screening systems.

Argolight's non photobleaching fluorescent sub-micrometer patterns are suited to fulfill the requirements of a reference tool for the calibration and alignment of fluorescence microscopes. They provide sub-resolution feature sizes of 250 nm in xy- and <1.5 µm in the z-direction, respectively; spectral features, including broad excitation and emission spectra, short lifetime, and various intensity levels; and long-term stability. In addition to fluorescence, the same patterns exhibit bright- and dark-field contrast, differential interference contrast (DIC), and phase contrast, which make them also suitable to calibrate these types of microscopes. Associated software facilitates not only performance monitoring, but also setup quality management. While the software does not currently fully automate the process, the process could be entirely automated if it was integrated into image-acquisition software—an idea we are proposing to microscope manufacturers.

REFERENCES

1. S. M. Reiss, "Quality and standards: Making bioimaging 'measure up,'" BioOptics World, 3, 1, 14–18 (Jan. 2010).

2. A. Dixon, Springer Ser. Fluoresc., 6, 3–24 (2008).

3. U. Resch-Genger et al., J. Fluoresc., 15, 337–362 (2005).

Arnaud Royon is chief technical officer and co-founder of Argolight SA, Talence, France; www.argolight.com.

Sponsored Recommendations

Request a quote: Micro 3D Printed Part or microArch micro-precision 3D printers

April 11, 2024
See the results for yourself! We'll print a benchmark part so that you can assess our quality. Just send us your file and we'll get to work.

Request a Micro 3D Printed Benchmark Part: Send us your file.

April 11, 2024
See the results for yourself! We'll print a benchmark part so that you can assess our quality. Just send us your file and we'll get to work.

Request a free Micro 3D Printed sample part

April 11, 2024
The best way to understand the part quality we can achieve is by seeing it first-hand. Request a free 3D printed high-precision sample part.

How to Tune Servo Systems: The Basics

April 10, 2024
Learn how to tune a servo system using frequency-based tools to meet system specifications by watching our webinar!

Voice your opinion!

To join the conversation, and become an exclusive member of Laser Focus World, create an account today!